Model CTR2000

Precision thermometer

Multi-channel version

Selected documents
Sales contact

WIKA Croatia d.o.o.

Tel.:
+385 1 6531-034
Fax:
+385 1 6531-357
info@wika.hr

Applications of precision thermometers

  • Precision thermometer for very accurate temperature measurements in a range of -200 ... +962 °C
    (-328 ... +1,763.6 °F)
  • Reference instrument for testing, adjusting and calibrating temperature measuring instruments in factories and calibration laboratories
  • Self-contained, complete system, also suitable for on-site measurements/calibrations

Special features of precision thermometers

  • Temperature range -200 ... +962 °C (-328 ... +1,763.6 °F)
  • Accuracy: 0.01 °C
  • 2 channels as standard with optional 8 channels
  • 4-wire measurement, optional 3-wire
  • SMART probe review and editor

Description of precision thermometers

The model CTR2000 precision thermometer comprises two measurement versions: One version accepts only 4-wire PRTs, the other version accepts both 3- and 4-wire PRTs.

The digital high-performance 2- or 8-channel, 3- and 4-wire precision thermometers are suitable for use with calibrated and uncalibrated Pt100 probes. It is characterised by its extended features for high usability and low life costs.

Using calibrated probes with the CTR2000 precision thermometer series choose between storing calibration data into memory in the instrument or into the probe’s proprietary SMART connector. Calibration data stays permanently with the probe in this connector, which the precision thermometer CTR2000 series instantly recognises, ensuring best measurement accuracy.

Using a built-in calibration routine, the CTR2000 instrument range can be recalibrated using a calibrated, traceable reference resistor. This feature is password protected.

Other features have been included for CTR2000 range when using SMART probes, like the temperature watchdog and the calibration status monitor.

The temperature watchdog continuously monitors the temperature of SMART probes and, if any are used outside their specified working range, the probe history record is updated.

With this way, it can be seen at a glance if the probes have ever been used outside their calibrated range, as this may affect their calibration validity. The calibration status monitor alerts with an on-screen message when a SMART probe is due for recalibration.

Precision thermometer CTR2000

Precision thermometer CTR2000

  • Precision thermometer CTR2000
Flyer

Advantages of AC resistance thermometry bridges

Wish list - Documents (0)

There are currently no documents on your wish list. You can add any document from the download area to your wish list. To do this, select the desired language using the flag symbol.

The maximum number of 20 entries was reached.

To receive your download link via e-mail, enter your address here:

The e-mail address is invalid.